Ozone Test Chamber Specification
Model | XB-OTS-80S | XB-OTS-150S | XB-OTS-225S | XB-OTS-408S | XB-OTS-800S | XB-OTS-1000S |
Inner chamber size W*H*D(cm) | 40*50*40 | 50*60*50 | 50*75*60 | 60*85*80 | 100*100*80 | 100*100*100 |
Exterior chamber size W*H*D(cm) | 90*90*150 | 95*150*105 | 95*165*115 | 105*175*135 | 145*190*135 | 145*190*155 |
Temperature Range | RT to 90°C | |||||
Humidity Range | 45%RH ~98%RH | |||||
Temp.&Humidity Fluctuation | ±0.2°C, ±0.5%RH | |||||
Temp.&Humidity Uniformity | ±1.5°C,±2.5% RH when RH is ≤ 75%,±4% RH when RH is > 75% (tested with the empty chamber, 30 minutes after stabilization) | |||||
Temp.& Humidity Resolution | 0.01°C,0.1%RH | |||||
Ozone Concentration | 0 to 1000 PPHM | |||||
Ozone Control Accuracy | ±10% | |||||
Ozone Generation | Static emissions | |||||
Sample Rotating Speed | 1 times per minute | |||||
Exterior Chamber Material | Stainless steel plate+Powder Coated | |||||
Interior Chamber Material | SUS#304 stainless steel plate | |||||
Insulation | PU & Fiberglass wool | |||||
Ozone Analyzer | Imported ozone density analyzer | |||||
Ozone Generator | Silent discharge type ozone generator | |||||
Wind Circulation | Sirocco fan | |||||
Heating System | SUS#304 stainless steel high-speed heater | |||||
Humidifying | Surface evaporating system | |||||
Cooling | Imported compressor,finned type evaporator,air-cooling condenser | |||||
Dehumidifying | ADP critical dew-point cooling/dehumidifying method | |||||
Controller | Digital electronic indicators+S.S.R.With PID automatic calculation capability | |||||
Other Components | Vacuum glass inspection window,cable port(50mm),control status indicator lights, chamber lamp,loading shelves (2pcs for free) | |||||
Safety Devices | Overheating circuit-breaker,compressor overload protection,control system overload protection, humidifying system overheating protection, breakdown indicator light | |||||
Power Supply | AC 1 phase 220V; 3 phase 380V 60/50Hz |
Ozone Chamber Features
-
External chamber: Arc design, electrostatically sprayed coat.
-
Inner chamber: SUS#304 stainless plate, anti-corrosion, and water resistance.
-
The door has a transparent window (220*320mm) for observation (lighting inside). It has the advantages of transparency, heat insulation, and explosion-proof.
-
The air duct of the windway is in the same direction as the specimen growth. A high-tensile silicone seal is used between the test door and body to ensure the test area is closed.
-
Cooling system: Installed at the bottom. The Tecumseh compressor is originally from France.
-
High-quality fixed wheels in the bottom, easy to move.
-
Heating system: completely independent, nickel-chromium alloy electric heater.
-
Heater control way: contact cycle pulse width modulated, SSR(solid state relays)
-
Output power calculus by microcomputer to control a smooth, reliable.