Standards of Formaldehyde Emission Test Chamber
Formaldehyde Emission Test Chamber is a formaldehyde emission detection and testing equipment developed according to the national standard GB18580-2017, LY/T1612-2004 American standard ASTMD6007-2, European standard EN717-1, international standard ISO16000-9, and German standard DIN EN 16516-2018.
- GB18580-2017 “Indoor Decoration and Decoration Materials – Formaldehyde Emission Limits in Artificial Panels and Their Products”
- GB/T39600-2021 “Classification of Formaldehyde Release from Wood-based Panels and Their Products”
- GB18587-2001 “Indoor Decoration and Decoration Materials – Limits for the Release of Harmful Substances from Carpets, Carpet Backings, and Carpet Adhesives”
- GB17657-2022 “Experimental Methods for Physical and Chemical Properties of Artificial Panels and Decorative Artificial Panels”
- EN 717-1 “Measurement of formaldehyde emissions from wood-based panels – Environmental chamber method”
- ASTM D6007-02 “Standard Test Method for Determination of Formaldehyde Concentration in Gases Released from Wood Products in a Small Scale Environmental Chamber”
- LY/T1612-2004 “1m Climate Box Device for Detection of Formaldehyde Release”
- GB50325-2020 “Code for Indoor Environmental Pollution Control of Civil Building Engineering”
- DIN EN 16516-2018 Building products – Assessment of hazardous substance release
Formaldehyde Emission Test Chamber Technical Parameter
Item | Technical Parameter |
---|---|
Model | XB-OTS-U1000 |
Inner chamber volume | (1±0.02)m3 |
Temperature range | (15—40)℃ (Temperature deviation±0.5℃) |
Humidity range | (30—80)%R.H. (adjusting accuracy:±1.5%R.H. ) |
Air exchange rate | (0.2-2.0) time/h (accuracy0.05 time/h), |
differential pressure | 10±5Pa Micro positive pressure, |
Air flow rate | (0.1—0.3)m/s (Continuously adjustable) |
Sampler suction speed | (0.25—2.5)L/min (adjusting accuracy:±5%) |
Sealing performance | When there is an overpressure of 1000Pa, the gas leakage is less than 10-3 × 1m3/min, with a gas flow difference of less than 1% between the inlet and outlet |
External dimensions | About (W1100×D1760×H1820)mm |
Voltage | 220V 16A 50/60HZ |
Power | Rated power: 3KW, operating power: 2KW |
Background concentration control | Formaldehyde concentration ≤ 0.006 mg/m ³ |
Adiabatic property | Climate chamber walls and doors should have effective thermal insulation |
noise | The noise value during the operation of the climate chamber shall not exceed 60dB |
Continuous working hours | The continuous working time of the climate chamber shall not be less than 40 days |
*Humidity control methods | The intelligent combination of dry wet method and dew point method is used to control the relative humidity of the working cabin. Taking advantage of the advantages of dry wet method and dew point method, the set humidity can be quickly reached within one hour through dry wet method. After reaching the set humidity, it automatically switches to dew point method to assist in humidity control, achieving very stable results. The humidity fluctuation is basically not more than 1.5%, far superior to the standard requirement of 3%. And it will not generate water droplets on the bulkhead; |
control system | The touch control screen is used as the dialogue interface for personnel to operate equipment. The Siemens PLC control system in Germany uses the touch control screen as the dialogue interface for personnel to operate equipment. It can directly set and display the temperature, relative humidity, temperature compensation, dew point compensation, dew point deviation, temperature deviation, working time inside the box, and can automatically record and draw control curves. |
Refrigeration system | We use ASPERA compressor imported from Italy. |
Security protection system | Refrigeration system high pressure and overcurrent protection, heater overtemperature protection, motor overcurrent protection, water shortage protection, air pump overheating, overcurrent, overpressure protection, etc. |
Features of Formaldehyde Emission Test Chamber
1. Inner Cavity and Structure: The inner cavity is made of 1.5 mm thick 304 mirror stainless steel, ensuring stability, durability, and strong sealing. It has a smooth surface preventing condensation and formaldehyde adsorption. The constant temperature box uses hard foam and silicone rubber for good insulation and sealing. Equipped with a forced air circulation device for uniform temperature and humidity.
2. User Interface: A 7-inch touchscreen provides an intuitive interface for setting and displaying temperature, humidity, compensation values, and deviations. It uses imported sensors to record and draw control curves, with dedicated software for system control, data display, playback, fault recording, and alarms.
3. Operational Stability: Uses industrial modules and imported programmable controllers for stability and reliability, ensuring long-term, trouble-free operation, reducing costs, and simplifying maintenance with fault self-inspection and prompts.
4. Optimized Control Program: The control program and interface are designed according to experimental standards, making operation simple and convenient.
5. Humidity Control: Uses intelligent integrated control of wet, dry, and dew point methods for stable humidity changes, improving control accuracy.
6. Temperature Sensor: Utilizes a high-precision platinum resistor as the temperature sensor, providing high accuracy and stable performance.
7. Heat Exchangers: Equipped with advanced heat exchangers for high efficiency and reduced temperature gradients.
8. Imported Components: Key components include imported compressors, Siemens modules, and industrial control PLC controllers, ensuring reliability.
9. Protection Devices: The box and water tank have high and low temperature and water level alarm protection measures.
10. Compact Structure: The integrated and compact structure makes installation, debugging, and use simple.
11. Convenient Features: Includes a makeup water pump for easy water replacement and cleaning, enhancing work efficiency and user convenience.
Working Conditions of Formaldehyde Emission Test Chamber
Environmental conditions
a) Temperature: 15-30 ℃;
b) Atmospheric pressure: 86-106kPa
c) There is no strong vibration around
d) There is no strong magnetic field influence around;
e) There is no high concentration of dust or corrosive substances around
Power supply conditions
a) Voltage: 220 ± 22V
b) Frequency: 50 ± 0.5Hz
c) Current: not less than 16A
Water supply conditions
a) Distilled water with a water temperature not exceeding 30 ℃
b) The placement position must ensure good ventilation and heat dissipation conditions (at least 0.5m away from the wall)